Ventilator Analyzer

The VA-01 is a cutting-edge ventilator analyzer designed for accurate calibration of ventilators, high-flow nasal cannula (HFNC), and CPAP/BiPAP systems.

Key Features

  • Measures bidirectional respiratory flow
  • Measures pressure inside the respiratory tubing
  • Measures oxygen concentration within the respiratory tubing
  • Measures temperature and humidity inside the respiratory tubing for gas measurement compensation
  • Measures atmospheric pressure for gas measurement compensation
  • Supports display of respiratory parameters based on seven standard gas measurement conditions: BTPS, STP20, STP21, STPD0, STPD20, STPD21 and ATP as well as user-defined environmental conditions
  • 5″ color touch screen for easy operation
  • IntVent software for use with a computer
  • Backup RAM for storing parameter data in the device
  • Built-in Li-ion battery provides approximately 4 hours of continuous operation
  • Compatible with ventilators, HFNC, and CPAP/BiPAP devices

Product Specifications


Pressure

Maximum applied pressure400 mbar
Range+/- 160 mbar
Resolution0.1 mbar
Span accuracy0.75% reading or 0.1 mbar, whichever is greater
Sampling rate200 Hz


Flow

Operating+/- 200 sLPM
Accuracy2.00 % reading or 0.1 sLPM whichever is greater
Resolution0.01 when flow < 100 sLPM
1.00 when flow > 100 sLPM
Sampling rate200 Hz


Volume

Range0.1 to 60 L
Accuracy2.00 % reading or 0.05 L whichever is greater
Resolution0.1 mL


Temperature and humidity

Temperature range10 to 40°C
Temperature accuracy+/- 0.5°C
Resolution0.1°C
Humidity range20 to 80% RH non condensing
Humidity range accuracy3% RH from 20 – 80% RH
Resolution1% RH


Oxygen concentration

Range0 to 100% vol.
Accuracy+/- 2% vol.
Resolution0.1% vol.
Inspiratory/Expiratory ratio (I : E) 1 : 600 – 600 : 1
Inspiratory tidal volume (VTI) 0 to 60 liters
Expiratory tidal volume (VTE) 0 to 60 liters
Inspiratory time (TI) 0 to 60 seconds
Expiratory time (TE) 0 to 60 seconds
Inspiratory peak pressure (PPEAK) +/- 160 mbar
Inspiratory pressure (PINS) +/- 160 mbar
Positive end expiratory pressure (PEEP) +/- 160 mbar
Inspiratory maximum pressure (PMAX) +/- 160 mbar
Minute volume (MV) 0 to 100.0 LPM
STP20Standard temperature 20°C, standard pressure
1013.25 hPa (760 mmHg), Actual humidity
STP21Standard temperature 21°C, standard pressure
1013.25 hPa (760 mmHg), Actual humidity
STPD20Standard temperature 20°C, standard pressure
1013.25 hPa (760 mmHg), dry (0% humidity)
STPD21Standard temperature 21°C, standard pressure
1013.25 hPa (760 mmHg), dry (0% humidity)
BTPSBody temperature 37°C, ambient pressure,
saturated (100% humidity)
STPD0Standard temperature 0°C, standard pressure
1013.25 hPa (760 mmHg), dry (0% humidity)
ATPActual temperature, pressure and humidity
Category Flow Analyzer only Total includes enclosure
Weight 1.1kg. 2.7kg.
Size (L x W x H) 30 x 17 x 8 cm. 41 x 35 x 14 cm.
  • Air, O₂
  • Operating temperature: 10 to 40°C
  • Storage temperature: 0 to 50°C
  • Operating relative humidity: 40 to 60 %RH
  • Input: 100 – 240 Vac, 50/60 Hz
  • Output: 12 Vdc, 1A
  • Battery: Rechargeable Li-ion, 2×4.2V, 2,600 mAH
  • Charging time: 8 hours from 0 to 100%
  • Operating time with battery: 4 hours maximum
  • USB A cable for communication with PC